In the case of air sampling, calibration (now more commonly referred to as flowrate verification) is the setting of the air flow through the sample medium to the recommended level for the method being used. The flow should be checked before and after each sample. SKC offers a range of calibration equipment, and the choice of which calibrator depends on:
A calibrator can either be to a primary or secondary standard. A Primary standard is an instrument that bases measurements on direct, measurable linear dimensions that will not change over time, or be altered by temperature or atmospheric pressure, such as the soap bubble flowmeter range. Secondary standard devices base their calibration upon primary standards and may be affected by temperature, atmospheric changes and mishandling. They can be used to obtain accurate, reliable measurements of pump flow rate if their calibration can be traced to a primary standard and they are handled with reasonable care.
SKC also offer a calibration service for your air sampling pumps.
https://www.skcltd.com/calibration.html; For further information please visit